Lachat Instruments: Laboratory FIA, IC and IL500 Series TOC/TN/P/N On-line Solutions
Contact Us | Links
Product Info | Components | Specifications | FAQ | Downloads | Request Info
IC+ Option
Components

Last update:
September 4, 2008


Eluent Pump
The high pressure pump delivers eluent to the IC manifold. The eluent pump incorporates a diaphragm damper which reduces pulsation in the system by as much as 98%. It includes an isolated pressure transducer which adds no volume to the system. The eluent pump has a keypad to set the flow rate and the pressure limit. The module stops pumping automatically if the pressure exceeds the pressure limit. The display on the pump shows the flow rate, maximum and minimum allowable pressures and the actual system pressure.

Sample Processing Module
A major component of the IC Option is the IC Sample Processing Module (ICSPM). A Sample Processing Module is also known as a channel. The SPM or channel is part of the System Unit. The ICSPM would be added to the FIA System Unit you currently own and consists of the following:

High Pressure Sample Injection Valve
Ion Profiling Module containing the guard and analytical columns
MicroSuppressor™
Conductivity Detector

© 2000-2008 Lachat Instruments  ::  Hach.com  ::  All rights reserved  ::  Legal Notice